Views: UV crosslinking is a key step in Southern blot, Northern blot, dot blot and EMSA workflows, fixing nucleic acids onto nylon or nitrocellulose membranes before hybridisation. Crosslinking quality is determined by the accumulated UV dose reaching the membrane, not by irradiation time. This guide explains how to choose between energy, time and preset modes, set a reproducible dose for 254 nm wet membranes, and run the seven-step procedure using the LUYOR UCL-3200 and UCL-3500 as examples. It also covers the reference dose table by experiment, troubleshooting for weak or excessive crosslinking, tube maintenance and calibration habits, so your conditions can be written into a repeatable protocol and reported clearly in publications.
Membrane crosslinking for Southern and Northern blot can look like pressing a start button, but what actually governs hybridisation signal and batch-to-batch repeatability is the accumulated UV dose received by the membrane surface. Drawing on the standard configuration of a UV crosslinker, this guide covers mode selection, dose setting, operating steps and common mistakes, with the LUYOR UCL-3200 and UCL-3500 as reference instruments whose parameters are listed on their model pages.
Set the target UV dose directly in mJ/cm2. A built-in UV sensor samples chamber intensity in real time and integrates automatically, stopping the lamp once the set dose is reached. As tubes age or supply voltage fluctuates, the instrument extends irradiation time to hold the end-point dose steady. Conditions remain comparable between batches and between instruments, which suits formal records and publication.
A fixed number of seconds or minutes is applied regardless of lamp output. After tube decay, the same 30 seconds delivers a much lower dose, weakening signal and destabilising results. Use it when reproducing an older protocol with a fixed duration; it is not suggested as a routine crosslinking method.
Many instruments ship with a built-in 120 mJ/cm2 standard crosslinking program for one-touch recall, and allow several common doses to be stored by the user.
Routine nucleic acid crosslinking uses 254 nm UV; 302 nm and 365 nm are used for mutagenesis and photoactive-molecule crosslinking and are not suited to standard Southern or Northern fixation. The doses below assume a wet membrane laid face-up.
| Experiment | Reference dose | Notes |
|---|---|---|
| Southern blot (DNA, nylon membrane, wet) | 120 mJ/cm2 | General reference used in many published protocols |
| Northern blot (RNA, nylon membrane, wet) | 120 mJ/cm2 | RNA over-crosslinks more readily; avoid increasing dose without reason |
| Dot blot, slot blot, colony / plaque lifts | 120 mJ/cm2 | General nylon membrane value |
| Nitrocellulose (NC) membrane | 80-100 mJ/cm2 | NC binds less strongly; higher doses can reduce signal |
| Small RNA such as miRNA and siRNA | 80-100 mJ/cm2 | Small RNA is more UV-sensitive; optimise with a gradient pilot |
| PCR UV decontamination, DNA mutagenesis | Set as needed | Sterilisation and mutagenesis are not membrane crosslinking; establish dose separately |
Wait for self-check to finish and confirm the 254 nm wavelength. Check that the chamber is clean, tubes show no blackening and the sensor window is unobstructed.
Press the Program / mode key, select Energy and confirm the display unit is mJ/cm2 or J/cm2.
For routine nucleic acid membranes enter 120 mJ/cm2 (0.12 J/cm2) and confirm with Enter; on preset models, recall the Auto-crosslink program directly.
After transfer, keep the membrane slightly damp with the nucleic acid face up toward the tubes. Lay it flat at the chamber centre without stacking; distribute multiple membranes evenly for uniform irradiation.
Close the chamber door and press Start. Do not open during irradiation, never look directly at the UV source and use UV protection.
The unit stops and beeps when the dose is reached. Record the set dose and the actual irradiation time; a longer time signals tube decay.
Move directly into pre-hybridisation after crosslinking and avoid prolonged exposure to strong light.
Under-crosslinking (dose too low): nucleic acid is not fixed firmly and detaches during hybridisation and washing, giving weak signal, absent bands and poor repeatability.
Over-crosslinking (dose too high): extensive UV damage to bases prevents probe binding and again lowers signal; samples subjected to repeated stripping and re-probing become harder to strip.
If the same dose takes markedly longer than before, the tube is decaying and should be assessed for replacement.
Wipe the sensor window with lint-free tissue; dust causes inaccurate readings and dose drift.
Long-wave UV does not effectively mediate crosslinking between nucleic acid and nylon membrane amines.
Use a professional UV radiometer for periodic calibration of the instrument.
Record wavelength plus dose in mJ/cm2 rather than seconds alone, so the work can be repeated and reviewed at submission.
With a fresh tube at high output, typically 25-35 seconds. As the tube ages, the same 120 mJ/cm2 may take 45-60 seconds. This is normal compensation in energy mode, not a fault.
Not suggested. Nylon tolerates higher UV and the reference is 120 mJ/cm2; NC is more fragile, so start testing at 80-100 mJ/cm2.
Irradiation time (s) = target dose (mJ/cm2) divided by real-time UV intensity (mW/cm2). The drawback is that intensity must be re-measured after every tube change or environmental shift, so long-term use is not suggested.
For short-term storage, sandwich the membrane in filter paper and keep it dark at room temperature. For longer storage, seal at 4 C to avoid nucleic acid degradation.
The LUYOR UCL-3200 UV crosslinker and LUYOR UCL-3500 UV crosslinker are 254 nm instruments with energy mode and a built-in UV sensor, suiting routine Southern and Northern blot work and PCR contamination control. Choose by membrane area and throughput; parameter comparison, selection guidance and stock enquiries are available.
For model comparisons and detailed specifications, visit the UV crosslinker category. For help matching a dose and configuration to your experiment, contact Xinysun Instruments (Shanghai) for selection advice.