Room-Temperature Single-Domain Liquid Crystal Elastomer in 30 Minutes: LUYOR-3109 Two-Stage Thiol-Ene Photoaddition Method

Updated:2026-11-02Views:

Reversible two-way shape memory in a liquid crystal elastomer requires the mesogens to line up across the whole sheet, that is, a monodomain; in a multidomain sample the local contraction directions cancel and the sheet barely moves. This article takes apart a two-stage thiol-ene photoaddition route: expose a dilute solution at room temperature, then stretch the film and irradiate both faces, reaching a monodomain sample in 30 minutes. It also explains where the parameters come from - 365 nm, 20 cm, 17.12 mW/cm² - and how the dose window is held.

Making a monodomain liquid crystal elastomer in 30 minutes: what the paper did

Liquid crystal elastomers (LCE) are materials that attach liquid crystal mesogens to a crosslinked rubber network. What makes them interesting: heat them above the nematic-to-isotropic transition and the material contracts along the mesogen alignment direction; cool it and the original length returns, which is reversible two-way shape memory. The precondition is that the mesogens line up across the whole sheet in one direction, a state called monodomain. In a multidomain sample the local contraction directions cancel and almost no motion is visible at the macroscopic level.

In 2017 the group of Yang Hong at Southeast University published a much faster route in Polymer Chemistry: two-stage thiol-ene photoaddition starting from liquid crystal monomers, at room temperature, in open air, one pot, producing a monodomain polysiloxane LCE film within 30 minutes. The methods section names two instruments - the UV source LUYOR-3109 benchtop photocatalytic LED UV lamp, and the LUYOR-340 UV radiometer used to measure intensity. The exposure part of this process is unpacked below.

ItemContent
Paper titleA room-temperature two-stage thiol–ene photoaddition approach towards monodomain liquid crystalline elastomers
Journal and yearPolymer Chemistry, 2017, 8(8), 1364–1370
DOI10.1039/C6PY02096H
InstitutionSchool of Chemistry and Chemical Engineering, Southeast University (Nanjing)
UV sourceLUYOR-3109, 365 nm (230 V / 0.20 A as recorded in the paper)
Intensity measurementLUYOR-340 UV radiometer, range 0–20 mW/cm², 290–390 nm
Key resultRoom temperature, open air, one pot and two steps, monomer to monodomain film within 30 minutes

Counting the time: where the conventional routes are slow

The value of the paper lies mainly in "fast" and "no need to exclude air". To see which problem it solved, look at the time cost of the main routes before it.

RouteMethodTimeMain barrier
Finkelmann two-step hydrosilylationPoly(hydrogensiloxane) + liquid crystal monomer + crosslinker, pre-crosslinked for several hours, mechanically stretched, then fully crosslinkedPre-crosslinking several hours, full crosslinking 1–2 daysPlatinum catalyst, slow reaction, inert atmosphere and temperature control needed
Yakacki thiol-acrylateSpontaneous thiol-acrylate Michael addition as pre-reaction, then photopolymerization to full crosslinking12 h pre-reaction + several minutes of photopolymerizationLong pre-reaction; the acrylate system is sensitive to oxygen
Liquid crystal cell methodFill a surface-treated cell with the monomer mixture, align, then photopolymerizeHours to daysCell assembly, filling, holding in the liquid crystal phase, opening the cell (some require HF etching of the glass)
Two-stage thiol-ene photoaddition (this paper)Partial photocrosslinking in dilute solution → stretching → suspended two-sided exposure to full crosslinkingStage one 5 minutes + stage two 15 minutesThe dose must be tight; stage one has to stay at partial crosslinking

The key difference is not how new the chemistry is, but that both crosslinking steps became light-initiated radical chain reactions, raising the reaction rate by orders of magnitude; at the same time the thiol-ene system tolerates oxygen far better than an acrylate system, which is why the whole operation can be done in open air.

Formulation and the two-stage flow at a glance

The paper gives the following standard formulation and procedure; all values are transcribed from the original.

ItemParameter
BackbonePMMS (poly(3-mercaptopropylmethylsiloxane)) 67 mg, containing 0.5 mmol thiol, taken as 100 mol%
Liquid crystal monomerMBB (4-methoxyphenyl 4-(1-butenyloxy)benzoate) 104 mg, 0.35 mmol, 70 mol%
CrosslinkerBUEB (1,4-bis(undec-10-enyloxy)benzene) 31 mg, 0.075 mmol, 15 mol%
PhotoinitiatorDMPA (2,2-dimethoxy-2-phenylacetophenone) 1.28 mg, 0.005 mmol, 1 mol%
SolventDichloromethane 0.5 mL; sonicate 2 minutes to remove bubbles
MoldPTFE rectangular mold, 2.0 cm long × 2.0 cm wide × 1.0 cm deep
Stage-one exposureRoom temperature, 5 minutes, lamp 20 cm above the liquid surface, intensity 17.12 mW/cm²
Demolding aidAdd a trace of hexane along the mold wall to help lift the film
Stretch alignmentCut into 2.0 cm × 0.8 cm strips, hang and stretch uniaxially to 130–140% of original length with about 6.85 g load
Stage-two exposureIn the suspended state, 7.5 minutes from each face, 15 minutes in total
ProductMonodomain LCE film about 500 μm thick

Why stage one has to be exposed in dilute solution at room temperature

This is the part of the paper that offers a great deal to an operator. Thiol-ene photoaddition is a radical chain reaction, and its speed is both the advantage and the trouble: the paper states plainly that control of intensity and exposure time is the key issue in "ensuring that the starting liquid crystal mixture undergoes partial crosslinking in stage one".

Their initial attempt was to spin-coat the liquid crystal mixture solution onto a PTFE plate as a thin film and expose it at 75 °C. The result: even with less than 1 minute of exposure the film crosslinked completely, hard enough that it could not be stretched, and no monodomain alignment was possible afterwards. Two things were changed:

  • Lower the reactant concentration: dilute to 0.5 mL of dichloromethane, so that the collision frequency for chain growth and chain transfer drops and the reaction slows;
  • Drop to room temperature: no more heating to 75 °C, one step lower in temperature.

With both changes in place, 5 minutes of exposure stops right at the partial crosslinking window - the film has formed, can be lifted and can be stretched, but the network is not locked. During those 5 minutes the dichloromethane also evaporates.

For comparison, the paper also made a multidomain sample: the same mixture was heated to 75 °C (about 73 °C above the melting point of the mixture) and UV-polymerized in bulk for 30 minutes to full crosslinking, for phase diagram measurement.

How the light source parameters were set: 365 nm, 20 cm, 17.12 mW/cm²

The methods section records the source as a hard citation, quoted here from the original:

"The mixture solution was UV-illuminated by using a UV lamp (230 V, 0.20 A, λ = 365 nm, LUYOR-3109, Shanghai LUYOR Instrument Co., Ltd) at room temperature for 5 min. The distance between the UV lamp and the surface of the solution was fixed at 20 cm where the light intensity was 17.12 mW cm⁻², measured by using a UV intensity meter (detection-range: 0–20 mW cm⁻², 290–390 nm, LUYOR-340, Shanghai LUYOR Instrument Co., Ltd)."

In operating language that is three things: wavelength 365 nm, lamp distance fixed at 20 cm, intensity measured at 17.12 mW/cm². Note that the third item was measured, not calculated - the paper used a UV radiometer at the workstation and held every sample at the same 20 cm, so that photon energy was consistent and comparable. This is a general requirement in photochemistry: change the lamp, the distance or the tube age and the dose changes, and results no longer line up.

Compared with the current catalogue specification: the LUYOR-3109 is rated at 9000 μW/cm² at 38 cm (that is 9 mW/cm²), with a 250 × 250 mm square uniform spot. The paper shortened the distance to 20 cm and measured 17.12 mW/cm², consistent with "closer means stronger". One caution: the paper records a 230 V / 0.20 A supply, while the supply and housing of the current shipping version should be taken from the product page; whichever version is used, intensity should be measured once at your own workstation rather than copying 17.12 mW/cm² from the paper.

Why this kind of process wants a 365 nm narrow-band cold source

Set the paper's conditions against the properties of the source and the process makes three hard demands.

Process requirementWhat the source has to provideCorresponding feature of the LUYOR-3109
The initiator must be excited effectivelyDMPA absorbs in the UV-A region; the main peak needs to sit at 365 nm9 × 365 nm high-power UV-A LEDs, narrow-band output, no visible or infrared spill
The reaction must run at room temperatureThe system must not be heated, or temperature will change both reaction rate and liquid crystal phaseCold source, no infrared output, housing around 50 ℃, does not bake the reaction system
20 minutes across two stages with comparable doseStable output, uniform spot, long continuous operationReady on switch-on, 24-hour continuous exposure, LED life about 50000 hours, square uniform spot keeps parallel samples at the same dose

The "no heating" point is easy to underestimate. The whole process works at room temperature because the source itself does not pour heat into the system: with a high-pressure mercury lamp, infrared would bake the solution in the mold to tens of degrees, the partial crosslinking window of stage one would drift with temperature, the evaporation rate of the solvent in the dilute solution would change, and the 5-minute time constant would no longer hold.

Stage two: stretch alignment and two-sided exposure

What stage one gives you is a partially crosslinked film with randomly arranged mesogens. To reach monodomain, the mesogens have to be combed into line:

  • Cut into strips of 2.0 cm × 0.8 cm, removing irregular edges;
  • Hang the strip and stretch it uniaxially with a load of about 6.85 g to 130–140% of original length;
  • While holding the stretch, expose each face for 7.5 minutes.

Exposing both faces rather than one has a direct reason: the film is about 500 μm thick and UV penetration into a polymer is limited, so a single-sided exposure creates a thickness gradient, well crosslinked near the lamp and under-crosslinked at the back; on cooling the internal stress is uneven, the film warps and its actuation is degraded. Equal exposure from both sides flattens the dose distribution through the thickness.

After stage two the network is fully locked and the mesogen alignment is fixed, which is what yields a monodomain film that contracts reversibly along the alignment direction on heating.

The three dose variables and one companion instrument

The paper's approach offers a reusable framework: dose = intensity × time, and intensity is set by lamp distance. In photoinitiated polymerization these three have to be managed together.

VariableValue in the paperOperating point
Lamp distanceFixed at 20 cmFix the same distance for every run and write it into the record; do not move the lamp by feel
Intensity17.12 mW/cm² measured at the workstationRe-measure after a lamp change, a position change or tube aging; use the radiometer, not the nominal figure
TimeStage one 5 minutes, stage two 7.5 minutes per faceUse a timer rather than visual judgement; the stage-one window is narrow, so shorter is safer than longer

The instrument used to measure intensity was the LUYOR-340 UV radiometer (range 0–20 mW/cm², 290–390 nm). In photochemistry and photopolymerization work this instrument plays the role of a balance - without it, any experience about exposure time is tied to one bench. To match a radiometer to a band and range, the model comparison in the UV radiometer section is a place to start.

How to verify the result: three criteria

Whether the film is monodomain and whether the transition temperature is suitable were checked in the paper with three criteria, and in practice the same three are used.

CriterionResult in the paperNote
DSC phase diagramTg = 1.5 ℃, Tiso = 49.5 ℃, room temperature inside the nematic window10 ℃/min under nitrogen, using the initial cooling and the second heating curve; glass transition below room temperature, clearing point under 50 ℃, suited to thermal actuation
1D wide-angle X-ray scatteringConfirms the mesophase is nematicResults in the main text, spectra in the supporting information
Polarized optical microscopy for monodomainA monodomain sample alternates bright and dark as the stage rotatesA multidomain sample keeps roughly constant brightness on rotation; a common quick check whether alignment succeeded

The control sample (bulk polymerization at 75 °C for 30 minutes, not stretched) is the multidomain reference: it has the same phase transition, but because the mesogen alignment is random, no reversible contraction along a single direction is visible at the macroscopic level.

An eight-step list that can be followed directly

StepOperationControl point
1Weigh PMMS, MBB, BUEB and DMPA to the stated ratio and dissolve in 0.5 mL dichloromethaneDMPA at 1 mol% should not be raised at will; more of it speeds up stage one and leads to over-crosslinking
2Sonicate for 2 minutes to remove bubblesBubbles leave defects in the film
3Pour into the PTFE mold (2.0 × 2.0 × 1.0 cm)The mold material must resist the solvent and release the film
4Fix the lamp distance at 20 cm and measure the workstation intensity with the radiometer, then record itIntensity is the reference value for later repeats
5Expose at room temperature for 5 minutes; the solvent evaporates meanwhileShorter is safer than longer; the film should lift off intact and still stretch
6Add a trace of hexane along the mold wall, lift the film, cut into 2.0 × 0.8 cm stripsLift slowly to avoid tearing the pre-crosslinked film
7Hang the strip, add about 6.85 g load, stretch to 130–140% of original lengthThe stretch ratio sets the degree of alignment and has to match your own system
8Hold the stretch and expose each face for 7.5 minutesEqual time on both faces; do not change the lamp distance when turning over

Common failure modes and remedies

SymptomLikely causeRemedy
Film too stiff to stretch after stage oneExposure too long or temperature too high, already fully crosslinkedShorten the time, return to room temperature, keep the solution dilute; reduce the initiator load if needed
Film will not lift and breaks on touchExposure too short, under-crosslinkedAdd 30 seconds per trial and record the threshold at which the film lifts intact
Same formulation, inconsistent results across batchesLamp distance changed, or lamp output has droppedFix the distance and measure intensity; keep a log of lamp hours
Two faces crosslinked to different degrees, film warps on coolingExposure was single-sidedExpose both faces for equal time and keep the same distance when turning over
Reaction much slower than in the paperVessel material absorbs 365 nm (ordinary glass attenuates noticeably)Switch to a quartz vessel, or measure the actual dose inside the vessel with a radiometer
Large variation between parallel samplesSamples stacked, or placed at the edge of the spotSingle layer inside the uniform spot, no stacking
Solvent evaporates too fast or too slowRoom temperature and ventilation differRecord ambient temperature and humidity; evaporation rate changes the stage-one concentration

Choosing among light sources in the same section

ModelForm and featuresSuitable duty
LUYOR-3109Benchtop, 365 nm, 9000 μW/cm² at 38 cm, 250 × 250 mm square uniform spot, continuous operationFixed workstation, full-area exposure, photopolymerization and photocatalysis that need long continuous runs
LUYOR-3410Handheld, can be mounted on a stand, 365 nm, 6000 μW/cm²Following the sample by hand, or local high-intensity exposure
Laboratory irradiation systemsIrradiation chambers and crosslinking equipment with quantitative dose settingQuantitative exposure where dose repeatability matters and shielding and interlock are required

If the reaction needs "a dose that is accurate and reproducible" rather than "a long exposure", chamber-type equipment is less demanding: distance, time and energy are managed by the instrument, and a change of operator does not easily shift the result. If you are simply setting up an open workstation for exploratory work, the benchtop LUYOR-3109 plus a radiometer is enough. When selecting, state four things clearly - the reaction system and volume, the intensity needed, the vessel material and the continuous working time - and the model and lamp distance can be checked together; see contact us.

Frequently asked questions

Is 365 nm necessary? Can 405 nm or a broad-band lamp be used?

It depends on the photoinitiator. DMPA absorbs in the UV-A region, which is why the paper used a 365 nm source. With an initiator that absorbs in the visible, the wavelength has to follow. The problem with a broad-band lamp is that the visible and infrared content contributes nothing to the reaction while bringing side reactions and heating, which works against both "room temperature" and "comparable dose".

Can the paper's 17.12 mW/cm² be copied directly?

No. That was a single reading taken with a radiometer at a 20 cm distance, tied to that lamp's output, that distance and that room. The right approach is to fix your own distance, measure once, and then calibrate exposure time against dose. Re-measure after changing a lamp or when output has clearly dropped.

Why is "open air" emphasized?

Many photopolymerizations (acrylate systems especially) are inhibited by oxygen: oxygen combines with chain-end radicals to form less reactive peroxy radicals, chain growth is held back, the surface stays tacky and crosslinking is incomplete, which is why nitrogen blanketing or a cover film is often used. A thiol-ene system has an extra chain-transfer step, and peroxy radicals can still abstract hydrogen from a thiol and keep the reaction going, so oxygen tolerance is much higher; that is the chemical basis for running this reaction in open air. The paper states that all operations were carried out at room temperature in open air.

How was the 5 minutes for stage one determined?

The paper says it was established through a large number of trials. The reusable idea is not to copy the number but to find the threshold at which the film lifts intact and still stretches, then leave a little margin; that time changes with formulation, solvent and lamp distance.

Can the film be thicker than 500 μm?

The paper reports about 500 μm. With a thicker film, UV attenuation through the film becomes the main consideration, the total two-sided exposure time may need to be extended, and cross-section inspection or mechanical testing should confirm that crosslinking is uniform through the thickness. If the gradient cannot be removed, warping during thermal actuation becomes likely.

Is a radiometer needed?

If results are to be compared across batches or across operators, yes. Exposure time is a proxy; what determines the extent of reaction is dose, and a radiometer converts "how long it was exposed" into "how much energy it received", which is what makes results comparable. The band has to cover the source (365 nm corresponds to a 290–390 nm probe).

Compared with a common handheld UV lamp, what does a benchtop source offer?

Spot uniformity and long-term stability. A handheld lamp is suited to local exposure that follows the sample, but with parallel samples it is hard to ensure that each one receives the same dose; a benchtop source with a square uniform spot puts the whole batch inside one spot, so dose consistency is much better, and it can run continuously without an operator swapping lamps.

The experimental conditions and data cited here come from Wang et al., Polymer Chemistry, 2017, 8, 1364–1370 (DOI 10.1039/C6PY02096H). To check a source model, lamp distance and dose plan against your own reaction system, send the system volume, target intensity, vessel material and continuous working time and we will work through the selection together - the contact page has the details. Related reading: Photocuring of microneedle patches: how to use the LUYOR-3109, laboratory UV lamp section.

Related sections and models

This article belongs to the laboratory research techniques section.

Other sections on the same product line include laboratory UV lamps (254 / 302 / 365 nm single-tube and dual-wavelength benchtop lamps) and life science instruments (the entry point for fluorescence imaging, UV crosslinking, UV transmission and cell irradiation).

Models mentioned in this article: LUYOR-3109, LUYOR-340, LUYOR-3410. Ranges, accuracy and accessory lists for each model are listed on the product page.

  • Official WeChat QR code

    Email: info@xinysun.com

  • Tel: 131-2093-6518

  • Address: 1525 Minqiang Road, Bldg 26, Songjiang, Shanghai

Official WeChat QR code Follow

Official WeChat QR code

Contact

131-2093-6518

Top