Views: Which lamp a photocleavable linker needs, and for how long, has a short answer worth stating first: 302 nm mid-wave or 365 nm long-wave suits the ortho-nitrobenzyl group, and the classic papers used a bench three-wavelength lamp of the XX-15 type. Irradiation runs from about 4 minutes to 30 minutes, set by the irradiance reaching the sample, the linker type and the sample form - in solution, on a solid support, or still on CPG. The note explains why 254 nm short-wave is not advised, since it induces thymine dimers, and works through the factors that stretch the time: distance, vessel material, sample form, and linker type with pH.
A PC linker (photocleavable linker) is a class of ortho-nitrobenzyl linking group, widely used for photocleavable biotin, light-controlled release tags and the affinity purification of synthetic oligonucleotides. "Which lamp, and for how long" are the two questions asked most often about this experiment. The answer up front: a 302 nm mid-wave or 365 nm long-wave UV lamp suits it - the classic literature used a bench three-wavelength lamp of the XX-15 type - and irradiation time runs from about 4 minutes to 30 minutes, set jointly by the irradiance of the lamp, the linker type and the sample form.
The absorption of ortho-nitrobenzyl PC groups falls in the near-UV (300-365 nm), and both lamp types have solid support:
The XX-15 laboratory tube UV lamp in our catalogue is a two-tube unit available at 254, 302 or 365 nm; changing the tube covers both photolysis requirements.
Cleavage time is set mainly by the irradiance reaching the sample surface. Reference values from the literature and supplier documentation:
In practice, irradiate for 5 minutes, take a sample and check cleavage efficiency by HPLC or a denaturing gel, then top up if needed; that is steadier than running the full time in one go.
254 nm deep UV will also cleave a PC group, but it induces thymine dimers in nucleic acids and damages the sample directly. Supplier guides state plainly that a stronger deep-UV source needs a 300 nm cut-off filter to protect the DNA. For photolysis, switch the lamp to 302 nm or 365 nm and leave 254 nm alone.
LUYOR XX-15/XX-20 laboratory tube UV lamp: two-tube construction, available at 254, 302 or 365 nm, the XX-15 rated 2×15 W and 1650 μW/cm² at 25 cm, an aluminium housing that can be hung, and a short-wave tube that can be flipped. PC linker photolysis, gel viewing and cutting, and bench sterilization are all covered by one unit. Users of the XX-15A and other models in the same series follow the same band choice and timing logic as this article.
Both bands cleave the ortho-nitrobenzyl group; the difference lies in efficiency and risk. 302 nm has higher photon energy and cleaves fast, releasing quantitatively from a solid phase in about 4 minutes in the literature, but it induces thymine dimers, so it suits short exposures where sample integrity is not critical. 365 nm is gentler and completes cleavage in about 10 minutes at 1 mW/cm², which suits samples that still have to go through enzymatic reactions, ligation or sequencing. If in doubt, start with 365 nm and switch to 302 nm only if efficiency falls short.
Irradiance falls roughly with the inverse square of distance, so the intensity at 25 cm is a quarter of that at 12.5 cm. The XX-15 is rated 1650 μW/cm² at 25 cm, and 10-25 cm is the usual working distance. For quantitative work, measure the irradiance at the sample with a UV radiometer and convert "how long" into "how much dose", so the conditions can be reproduced in another laboratory.
302 nm mid-wave is absorbed noticeably by ordinary glass and most plastics, so use a quartz dish, a UV-transmitting petri dish, or an open small-well plate. 365 nm is easier on the vessel and ordinary clear plastic is acceptable for short exposures, but for long exposures move to a UV-transmitting material, otherwise the dose that actually reaches the sample is discounted.
Sample and run HPLC or a denaturing polyacrylamide gel, then check that the starting-material peak and band have gone and that the released product is present. In practice, irradiate for 5 minutes, sample, and top up for another 5-10 minutes if needed; that is more controllable than running the full time at once. Note as well that an oligonucleotide still on the CPG support photolyses at only 5-10%, so cleave it off first.
Keep them dark at once: transfer to an amber tube or wrap in aluminium foil so that handling and bench lighting do not expose them again. Release of 3'-phosphate oligonucleotides depends on pH, being higher at pH 9.4 than at pH 7.4, so prepare the buffer for the downstream reaction. Use photolysis products within a short period rather than storing them long term.
For small volumes over short times it can serve as a stopgap, but a handheld source has unstable irradiance and an uneven spot, so repeatability is poor, and the intensity changes the moment the distance changes, which means the timing cannot be fixed. For quantitative work, batch processing or anything written into a method, a tube lamp such as the XX-15 / XX-15A at a fixed working distance, with a radiometer to record irradiance, is more reliable.
This article belongs to the frequently asked questions category.
Other categories on the same product line are UV crosslinkers (dose set in J/cm², the figure that matters for both nucleic acid crosslinking and material curing), laboratory UV lamps (single-tube and dual-wavelength bench lamps at 254 / 302 / 365 nm) and life science instruments (the single entry point for fluorescence imaging, UV crosslinking, UV transmission and cell irradiation).
Models mentioned: UCL-3200, UCL-3500, UVP CL-1000. Range, accuracy and accessory lists for each model are given on the product pages.
Email: info@xinysun.com
Tel: 131-2093-6518
Address: 1525 Minqiang Road, Bldg 26, Songjiang, Shanghai
Copyright 2026 Xinysun Instrument (Shanghai) Co., Ltd. All rights reserved.
ICP:ICP-2026046447