Breadcrumb: Home > Applications > Lab Research
Updated:2026-11-02
Views: In April 2025, a study from China Pharmaceutical University and Shanghai Junyu on gardenia extract against UVB photodamage appeared in Cosmetics. The exposure source recorded in the method section was a 302 nm UV crosslinker, which in the LUYOR product line is the UCL-3200M. This article unpacks the exposure scheme: why one study needed both 0.01 J/cm² for HaCaT cells and 0.4 J/cm² for a 3D epidermis model, why the low-dose end requires energy mode rather than timing, three operating details that decide whether the run works, and how the UCL-3200M and UCL-3500M divide the work. The recorded doses are 10 mJ/cm² and 400 mJ/cm², a fortyfold span.
In April 2025, the Laboratory of Skin Health and Cosmetic R&D Evaluation at China Pharmaceutical University, working with Shanghai Junyu Biotechnology, published a study in MDPI Cosmetics on gardenia fruit extract (Gardeniae Fructus extract, GFE) against UVB photodamage (Cosmetics 2025, 12(2), 72, open access). The method section records the exposure source as "UV crosslinker, Luyor (Shanghai)" at 302 nm - which in the LUYOR product line is the UCL-3200M .
From an instrumentation point of view, one feature of this study is worth unpacking: the same study used two doses - 10 mJ/cm² for the cell model and 400 mJ/cm² for the 3D epidermis model, a fortyfold difference. For one crosslinker to be accurate at both the 0.01 J/cm² end and the 0.4 J/cm² end, what matters is not "how many seconds" but the energy integral. This article reconstructs the exposure scheme, the dose logic and the operating points.
The two doses correspond to two models, and the difference lies in how many layers light crosses before reaching living cells:
In other words, the dose difference is not arbitrary; it is derived backwards from the barrier structure of the model. That is practical for anyone doing efficacy evaluation: change the model and the dose has to be reset, and copying numbers from the literature often wastes a run.
| Item | Cell model (HaCaT) | 3D epidermis model |
|---|---|---|
| Exposure source | LUYOR UV crosslinker (Luyor, 302 nm) | Same UVB source (the model is not restated in that section) |
| Wavelength | 302 nm (UVB) | 302 nm (UVB) |
| Dose | 10 mJ/cm² (0.01 J/cm²) | 400 mJ/cm² (0.4 J/cm²) |
| Medium during exposure | Thin PBS layer over the cells | Submerged in culture medium |
| Post-exposure handling | Fresh medium, culture continued for 24 h | GFE 50 μg/mL for 24 h, then readout |
| Readouts | ROS, T-SOD, CAT, GSH-Px, MDA; IL-6, IL-1β | FLG, LOR, IVL immunofluorescence (barrier proteins) |
The paper gives the dose but not the exposure time or working distance. That is precisely the point: the authors recorded a dose, not a time. A crosslinker is a closed chamber and the samples sit at fixed positions inside it, so there is no adjustable "lamp distance" to report; dose is the number that genuinely belongs in the method section.
UCL-3200M has an energy range of 0.000-9.999 J/cm², and 10 mJ/cm² converts to 0.010 J/cm², one thousandth of the range. That position is sensitive to the control method:
That is what energy mode is for: the microprocessor monitors UV intensity in the chamber during exposure and integrates over time, stopping as soon as the set dose accumulates. If the lamp dims it exposes longer; if intensity is high it stops earlier - what goes into the record is "10 mJ/cm²", a figure that can be checked, rather than "exposed it for a moment".
400 mJ/cm² is 40 times the cell end, and the exposure moves from seconds to tens of seconds or minutes. On the face of it, time mode would work, but the same problems appear: 3D model batches are run across different days, lamp condition and chamber temperature change, and a time-based protocol does not line up between batches - which shows up as vague complaints such as "the FLG staining came out faint in this batch".
The chamber is 32×26×15 cm, so a full 12-well Transwell plate fits flat and is exposed in one pass, which reduces within-batch variation. Another advantage of energy mode at the high-dose end is that no conversion is needed: whether the target is 10 or 400 mJ/cm², the setting is the same physical quantity, and changing model or project does not mean feeling out the time again.
The composition of GFE was characterized by UPLC-MS/MS: 10 iridoids were identified, with geniposide prominent among them, alongside genipin-1-gentiobioside, gardenoside acid, shanzhiside methyl ester, deacetylasperulosidic acid and asperuloside. Extraction used 20 volumes of water at 80°C, twice for 2 hours each, followed by activated carbon decolorization, macroporous resin purification, ethanol elution and freeze-drying.
The readouts fall into four layers, from the surface inward:
| Layer | Marker | Change in the UVB model | Effect of GFE |
|---|---|---|---|
| Oxidative stress | ROS (DCFH-DA) | Increased | Significantly suppressed at 12.5/25/50 μg/mL |
| Antioxidant enzymes | T-SOD, CAT, GSH-Px | Decreased | Recovered effectively |
| Lipid peroxidation | MDA | Increased | Reduced |
| Inflammatory cytokines | IL-6, IL-1β (ELISA) | Secretion increased | Significantly suppressed |
| Barrier proteins | FLG, LOR, IVL (3D model immunofluorescence) | Expression downregulated | Upregulated after 50 μg/mL for 24 h |
The paper cross-validated with proteomics and network pharmacology, and the numbers are worth noting:
The authors also state clearly in their conclusion that these mechanisms are predictions and leads that still require experimental verification. When writing efficacy claims, treating them as directions for later validation rather than established conclusions is the safer course.
The same series also includes a 302 nm UCL-3500M , and the site already carries an application article on it at a 50 mJ/cm² dose, on retinol and DNA repair (see the article on 302 nm UVB quantitative exposure ). Both can deliver quantitative UVB exposure; the choice depends on sample volume:
| Item | UCL-3200M | UCL-3500M |
|---|---|---|
| Lamp configuration | 6 × 8W | 6 × 15W |
| Chamber size | 32×26×15 cm | 44×34×15 cm |
| Energy range | 0.000-9.999 J/cm², 9 presets | 0.000-9.999 J/cm², 9 presets |
| Suited to | One plate or a few samples, a wide dose span, fine control at the low-dose end | Multiple plates in parallel, large models, batch exposure |
| Corresponding case | This article (10 and 400 mJ/cm²) | 50 mJ/cm² skin photodamage and DNA repair |
If the project runs mainly on 96-well plates and just one or two plates per batch, the 3200M chamber is a good fit; if a dozen plates or large ex vivo skin samples are exposed at once, the 3500M has more room. The wavelength version system is the same for both: M is 302 nm (UVB), the standard version is 254 nm (UV-C), and L is 365 nm (UV-A), so a version change keeps the same control logic and preset method.
The method section says "UV crosslinker, Luyor (Shanghai)" plus a 302 nm wavelength, without a model number. The LUYOR series is split by wavelength into three versions: 254 nm is the standard version, 302 nm is M and 365 nm is L. Matching 302 nm gives the UCL-3200M. If your own paper needs a model number, state the wavelength and the dose as well - those two are what make the work reproducible.
The UCL-3200M energy range is 0.000-9.999 J/cm², so 10 mJ/cm² is 0.010 J/cm², resolved to 1 mJ/cm². The key at the low-dose end is to avoid time mode: the exposure lasts a few seconds and manual timing carries a large relative error. In energy mode the sensor integrates to the set dose and stops, which is what makes the dose stable.
Because of the added stratum corneum. Two-dimensional HaCaT monolayers have a thin PBS layer over them and UVB reaches the cells almost directly; a reconstructed human epidermis has a stratum corneum and a stratified structure, and UVB is markedly attenuated before it reaches the living cell layer. The dose has to be reset to the barrier structure of the model, not copied.
It depends on the model. The paper kept a thin PBS layer on the cell model to prevent drying, and submerged the 3D model in culture medium. Either way, the liquid volume must be the same across a batch - liquid thickness directly changes the dose reaching the sample. Replace the medium after exposure, which also clears reactive oxygen species generated by medium photochemistry.
Not as a direct substitute. 254 nm is UV-C, with stronger DNA absorption and a different damage mechanism and penetration depth, so results cannot be compared with UVB literature. A UVB photodamage model needs the 302 nm version (UCL-3200M); the 254 nm version suits nucleic acid membrane crosslinking, PCR decontamination and UV mutagenesis, as described in the article on UV mutation breeding.
Literature: Zong K, Li X, Zhou F, Dong J, Huang Q, Wu J. Photoprotective Effect and Potential Mechanisms of Gardeniae Fructus Extract in UVB-Irradiated HaCaT Cells. Cosmetics 2025, 12(2), 72. DOI: 10.3390/cosmetics12020072 (CC BY open access). All data in this article are taken from that paper; for specific experimental conditions, refer to the original paper and the instrument manual.
This article belongs to the laboratory research techniques section.
Other sections on the same product line include UV transilluminators (gel and thin-layer chromatography excited from below, so bands can be read without turning the gel over), UV crosslinkers (dose set in J/cm², the figure that matters for both nucleic acid crosslinking and materials curing), and life science instruments (the entry point for fluorescence imaging, UV crosslinking, UV transmission and cell irradiation).
Models mentioned in this article: UCL-3200, LUV-450, XEPU-1126. Range, accuracy and accessory lists for each model are given on its product page.
Email: info@xinysun.com
Tel: 131-2093-6518
Address: 1525 Minqiang Road, Bldg 26, Songjiang, Shanghai
Copyright 2026 Xinysun Instrument (Shanghai) Co., Ltd. All rights reserved.
ICP:ICP-2026046447