Breadcrumb Home > Products > Industrial Test Instruments > UV Radiometers

Spectro-UV XP4000 UV-A / White / Blue Light Radiometer
  • Spectro-UV XP4000 UV-A / White / Blue Light Radiometer

Spectro-UV XP4000 UV-A / White / Blue Light Radiometer

OverviewOverview

XP4000 AccuPRO Plus radiometer reads UV-A 0-100 mW/cm2, white light 0-5382 lux and 450 nm blue light on one 3-in-1 sensor. Better than 5%, NIST traceable.

Model:
Updated:2026-11-26
Manufacturer:
Reference price:Request a Quote
Views:
DetailsSend inquiry

A UV radiometer measures UV-A irradiance on the test surface in μW/cm² or mW/cm² or, for visible light, illuminance in lux. In fluorescent penetrant (PT) and fluorescent magnetic particle (MT) inspection it settles one question before every acceptance call: is the black light still above the specified irradiance, and is the ambient white light still below its limit. XP4000 from Spectro-UV (AccuPRO Plus) reads UV-A, white light and 450 nm blue light on a single three-channel sensor, so a shop that also records blue-light output does not need a second instrument.

3-in-1 sensorUV-A, white light and 450 nm blue in one head
UV-A 0-100mW/cm², 320-400 nm
White 5382lux (0-500 fc), 460-675 nm
Blue 450 nm0-100 mW/cm², 410-475 nm
Better than ±5%NIST traceable calibration
In stock in Shanghaispec sheets and selection help on request

Why a radiometer belongs at the inspection station

Detectability in fluorescent PT and MT depends half on the process and half on the actual state of the light source. The intensity printed on a lamp is a factory figure. LEDs age, filters collect penetrant mist, drivers drift, and after a year of production use the irradiance reaching the part can sit well below what the acceptance document assumes. The eye cannot see any of this: the lamp still looks bright.

Inspection standards therefore ask for numbers rather than impressions: UV-A irradiance on the test surface above a stated minimum, ambient white light below a stated maximum. Both are measured at a fixed distance, at fixed points, with an instrument whose calibration can be traced to a national standard. That is the whole job of the radiometer: it turns “is the lamp still good?” into a reading that can go into a quality record.

What the instrument does not do. A radiometer does not change the light. Its value comes from the trend line built over months: same lamp, same distance, same points. A single reading says what the lamp is doing today; a series of readings says when to replace it.

What the three channels measure

ChannelMeasured quantityRangeSpectral rangeTypical use
UV-AUV irradiance (black light intensity)0-100 mW/cm²320-400 nmChecking that a PT or MT black light still meets the specified irradiance
VisibleWhite light illuminance0-5,382 lux (0-500 fc)460-675 nmChecking that darkroom white light stays below the limit, so fluorescence contrast survives
BlueBlue irradiance0-100 mW/cm²410-475 nmRecording blue light output for processes excited around 450 nm

The three channels share one sensor and are selected with the panel button, so every reading refers to the same position and the same angle. Reading UV-A with one probe and white light with another is a quiet source of disagreement between shifts, and it disappears when both numbers come from one head.

What makes a reading trustworthy

ParameterXP4000 specificationWhat it means on site
Overall accuracyBetter than ±5%, NIST traceableThe margin between pass and fail is far wider than the uncertainty of the instrument
Temperature coefficient+0.025%/°C over 0-50°CShop-floor temperature swings barely move the reading
Linearity±0.2%; the sensor sends its linearity correction to the readout at power-upReadings stay comparable across ranges, which is what a trend record needs
Filtering and responseBandpass interference filters, good cosine response, optical stack assembled in a Class 100 benchAngle changes do not distort the number, and unit-to-unit variation is small
Display4-digit auto-ranging LCD, 128×64, 2.25 in, backlit, 2 Hz refreshReadable in a darkroom, and no manual range or unit conversion

The instrument is an optical product: avoid drops, condensation and strong electromagnetic fields, keep fingers off the sensor window and clean it with lens tissue. Have it recalibrated on a schedule rather than after a reading looks wrong.

Building a lamp verification routine

Five steps turn a radiometer into a management system rather than an occasional check:

  • Fix the geometry. Write the distance (38 cm, or whatever the acceptance document states), the angle, the lamp setting and the measurement points into the work instruction.

  • Record a baseline. Measure a new or freshly calibrated lamp and keep the numbers as the reference for that station.

  • Recheck on a schedule. Per shift, weekly or quarterly, depending on the standard and your own quality plan. Recheck after replacing LEDs, changing a filter, or dropping the lamp.

  • Act on the trend, not on failure. When UV-A approaches the specified minimum, or white light drifts up, clean the filter, replace the lamp or fix the darkroom before detectability suffers.

  • Keep the certificate with the record. A reading without traceability is not evidence during a customer audit.

Frequently asked questions

Is a UV radiometer the same as a UV energy meter?

No. A radiometer reads instantaneous irradiance in μW/cm² or mW/cm² and answers “how strong is the lamp right now”. A UV energy meter integrates dose over time in J/cm² or mJ/cm² and answers “how much energy did the surface receive”. Lamp acceptance in PT and MT uses irradiance; curing and crosslinking processes use dose.

What is the difference between XP4000 and XP2000?

XP2000 (AccuPRO) has two channels, UV-A and white light. XP4000 (AccuPRO Plus) adds a 450 nm blue channel and uses a three-in-one sensor. If your process only needs UV-A and white light, the two-channel version is enough; if you record blue light output, choose XP4000.

Why measure blue light at all?

Some fluorescent inspection processes excite with blue light around 450 nm and read through yellow filter glasses. Those processes need the blue irradiance verified, and where the light output has to be documented, they need a separate blue channel. For UV-A and white light only, the blue channel is redundant.

How do mW/cm² and μW/cm² relate?

1 mW/cm² equals 1000 μW/cm². The instrument switches range automatically, showing μW/cm² below 10,000 and mW/cm² above it, with the unit shown next to the number. No manual conversion is needed.

Does the XP4000 meet ASTM E3022?

XP4000 is stated against ASTM E2297 and CEN/TR 16638 with NIST traceable calibration. ASTM E3022 is the dedicated standard for UV-A lamps and their meters in fluorescent PT and MT; if your acceptance document names E3022, check the exact clauses with us before ordering so the meter and the document match.

How is it powered, and what is in the case?

Four AAA NiMH rechargeable cells, charged internally from the supplied AC adapter, give about six hours of continuous scanning with roughly eight hours to a full charge. The standard set contains the AccuPRO Plus readout, the 3-in-1 sensor, the charger, a padded carrying case and a 0.9 m sensor cable.

Which plug version should I order?

XP-4000 uses US Type A/B, XP-4000/F European Type C/F, XP-4000/FA Australian and Chinese Type I at 220-240 V, XP-4000/FB British Type G. Shipments to mainland China normally take XP-4000/FA.

How often should the meter be calibrated?

Every 6 to 12 months in normal use, sooner after a drop, a long idle period or any reading that looks wrong. Calibration should be done by an accredited laboratory and comes with a certificate stating the traceability chain and the measured deviation. We can arrange the send-in for you.

A radiometer only pays for itself when the readings are comparable over time, which means one instrument, one geometry and one record. Compare the three meters on the UV radiometer range page, or look at the XRP-3000 if you need interchangeable UV-C, UV-B and luminance sensors, and the Labino Apollo 3.0 if the probe has to sit at a fixed point while you read from up to 5 m away. Further reading: how to choose a UV radiometer and ten questions on inspection lamps.

Send inquiry

Message

  • Product Name:

  • Company:

  • Your Name:

  • Phone:

  • Email:

  • Province / Region:

  • Address:

  • Message:

Copyright 2026 Xinysun Instrument (Shanghai) Co., Ltd. All rights reserved.
ICP:ICP-2026046447

SiteMap | Privacy Policy | FAQ

Official WeChat QR code

WeChat QR
Official WeChat QR code Follow

Official WeChat QR code

Contact Contact

+86-131-2093-6518

Back to top Top