Start here: what this page answers

These are the questions that come in before an order, grouped by instrument family. Each answer is short and links to the page that carries the full method, the numbers and the model comparison. If your question is about a specific acceptance document, send us the clause and we will answer against it rather than against a generic figure.

Black lights for fluorescent penetrant and magnetic particle inspection

How much UV-A irradiance does a black light need on the part?

Fluorescent penetrant and magnetic particle work calls for at least 1000 μW/cm² on the test surface in a darkened area. Quoting a lamp at 3000 μW/cm² or more leaves margin for the decline that comes with lamp hours, because a new lamp that passes today may not pass in two years. Ten questions on inspection lamps

365 nm or 395 nm, and why does it matter for documentation?

365 nm sits at the short end of UV-A: it excites penetrants, leak dyes and riboflavin efficiently and leaves little visible violet, so fluorescence contrast is high. 395 nm costs less but contains much more visible violet, and surface reflections are easily mistaken for indications. When photographs, cleanroom acceptance or audit records are involved, 365 nm is the wavelength to specify. 365 nm black light range

Flashlight, head-mounted or hanging lamp: which form fits the job?

Head-mounted for inspection that involves climbing or crawling, because it is hands-free and follows the line of sight. Flashlight for scanning large areas and patrol work. Hanging or bench models for fixed stations and batch acceptance, where continuous running matters more than battery life. How to use a UV inspection lamp

How do I tell whether white light leakage is too high?

In a darkened area, point the lamp at white paper or a fluorescent screen and look for an obvious violet-white patch. The measurable version is to read the white light on the workpiece with a radiometer that has a visible channel and keep it under the limit your procedure states, commonly 20 lux. Filter ageing, LED batch variation and a failed housing seal all raise leakage. Six questions on lamp compliance

What happens to an LED black light as it ages?

The usual symptom is a slow loss of intensity, not a sudden failure: after 1000 hours of accumulated operation output can fall by 10% to 30%, so a lamp that measured just over the threshold can drift into non-compliance without anyone noticing. If you find yourself holding the lamp closer to see fluorescence, that is the cue to measure it. Five common mistakes in UV-A inspection

Can a black light be used for leak detection as well as inspection?

The lamps overlap but the dyes and the working conditions differ. Leak detection adds a fluorescent tracer to the oil or water circuit and reads the seepage on the outside of the machine, which needs a strong 365 nm beam rather than a wide inspection field. Fluorescent leak detection guide

Does a black light need an external filter?

It depends on the emitter. Lamps built around 365 nm LEDs with a well-matched filter glass keep visible leakage low without a bolt-on filter; cheaper emitters rely on the filter to block the violet tail, and that filter degrades with heat and cleaning. What matters in the end is the measured white light figure, not the presence of a filter. Using a black light through the penetrant process

UV radiometers and lamp verification

Is a UV radiometer the same as a UV energy meter?

No. A radiometer reads instantaneous irradiance in μW/cm² or mW/cm² and answers how strong the lamp is right now. An energy meter integrates dose in J/cm² and answers how much energy the surface received. Lamp acceptance needs irradiance; curing and crosslinking need dose. UV radiometer range

Which matters more, more functions or traceable calibration?

Traceability. A radiometer exists to prove a lamp still meets a specification, and an auditor asks for the calibration certificate, not the feature list. A NIST traceable or ISO/IEC 17025 certificate is what makes the reading evidence rather than an opinion. How to choose a UV radiometer

How do I choose between the three meters?

By the measurement points and how you want to read. The XP4000 puts UV-A, white light and 450 nm blue on one cabled three-in-one sensor. The XRP-3000 uses a body that also accepts UV-C, UV-B, blue and luminance sensors. The Labino Apollo 3.0 puts a wireless probe on the point and lets you read from up to about 5 m away.

How often should it be calibrated, and what if it is dropped?

Every 6 to 12 months in normal use, and immediately after a drop, a long idle period or any reading that looks wrong. Send it to an accredited laboratory and keep the certificate with the lamp records, otherwise historical readings cannot be defended. Apollo 3.0 calibration and warranty

Why fix the distance every time?

Irradiance falls with distance, so a reading taken at 30 cm and one at 38 cm are not the same measurement. Writing the distance, angle, lamp setting and points into the work instruction is what turns a series of readings into a trend that can be acted on. Measurement geometry and traceability

Surface inspection and particle detection

White, yellow-green or UV: how do I choose the light for surface inspection?

Yellow-green around 520 to 555 nm sits where the eye is most sensitive, so it reveals scratches, oil film and fibres at low output without glare. White light is for colour judgement and general checks. UV is for fluorescence: residue, penetrant, adhesive, biofilm. 18 questions on surface inspection lights

Do I need both a black light and a surface inspection light for particle work?

They answer different questions. The black light shows fluorescence from organic residue and penetrant; the yellow-green or white lamp shows particles and contamination on the surface. A cleanroom programme usually carries both and assigns each to a step. Particle detection lamp selection

Can I inspect carrier inner walls and FOUPs with a flashlight?

Yes, and it is one of the more useful applications: the inside of a carrier is where particles collect and where a normal wipe test struggles to reach. A UV flashlight with a narrow head gets into the corners. Particle contamination in wafer carriers and FOUPs

What does an inspection light need for a coating line?

Reach and contrast: a hand-held yellow-green lamp at a shallow angle shows oil stains and fibres before paint goes on, which is cheaper than finding them after curing. Distance and angle matter more than raw brightness. Surface cleanliness before coating

Is a bright lamp enough, or do I need to measure it?

Measure it. Inspection lights lose output with hours and filter contamination, and the operator cannot see the loss. A periodic check at a fixed distance is the only way to know the lamp still does what the procedure assumes. Surface inspection lamp range

Stroboscopes, tachometers and sensors

Xenon or LED strobe: which one freezes motion better?

It comes down to flash duration rather than peak brightness. A short flash is what stops a fast surface without blur; a long flash leaves a smear no matter how bright it is. LED strobes have closed much of the old gap and are the practical choice for most production lines. Xenon or LED stroboscope

How do I read rpm on a shaft when there is no room for tape?

Either mark the shaft with a paint pen or use a sensor that does not need a target. Reflective tape needs a flat, clean, visible patch; on small or oily shafts a contrasting mark, a keyway or a coupling bolt often works better. Bench tachometry above 10,000 rpm

Laser or photoelectric sensor: what actually differs?

Beam size and standoff distance. A laser holds a small spot at a long distance, which suits a target you cannot get close to; an LED photoelectric head is more forgiving of angle and cheaper, and works well at short range on a clean target. Laser or photoelectric tachometer sensor

How do I wire a sensor into a PLC?

Establish three things first: NPN or PNP output, TTL or open collector, and the supply voltage the sensor expects. Getting the output type wrong is the usual reason a sensor appears dead on a PLC input. Wiring and output: NPN, PNP and TTL

The reading jumps between two values. Is the meter broken?

Usually it is reading a second target: a keyway, a coupling bolt, a label edge, or a reflection from a polished surface. Cover the extra target, change the angle slightly, or narrow the spot, and the doubling disappears. Sensor repair and troubleshooting

Can a strobe be used as a tachometer, or the other way round?

A strobe can measure speed by tuning until a mark stands still, which is useful when you cannot touch the machine, but it is a manual process and can be fooled at multiples of the true rate. A tachometer with a sensor reads directly. Most maintenance teams keep both. Stroboscope range and laser tachometer range

Reflective tape

Where should the tape go on a shaft?

On a clean, flat, visible face, perpendicular to the beam, with the whole patch inside the sensor spot. Avoid a seam, a radius or a surface that flexes; a patch that lifts at one edge gives an intermittent count rather than a wrong number. Reflective tape range

How big should the patch be?

Sized to the optical spot at your working distance, not to the shaft. A patch much larger than the spot is wasteful, and a patch much smaller returns too little light at long standoff. Pre-cut dots and squares remove the guesswork. Marking the shaft

Does tape work on a hot or oily machine?

Standard tape does not last long on either. High-temperature tape and a solvent wipe before application extend life, and on a hot surface the adhesive needs to be rated above the surface temperature, not the ambient. Tape grades

Can we buy tape printed with our own brand?

Yes. Glass-bead retroreflective face, several adhesive grades, widths from 10 to 80 mm, die-cut shapes matched to your optical spot, and packing counted per instrument. Samples are free, freight on the buyer's account. Reflective tape OEM for tachometer and balancer makers

UV crosslinkers and transilluminators

254, 302 or 365 nm for crosslinking?

254 nm is the standard for nucleic acid membrane crosslinking and has the highest efficiency. 302 nm penetrates a little better for thicker membranes. 365 nm has lower energy and is used with photosensitive dyes, or where DNA damage must be avoided. Choosing the crosslinker wavelength

Should I set seconds or joules?

Joules whenever the instrument supports it. Lamps decline with use, so a fixed 30 seconds on a new lamp and on a three-year-old lamp are different doses with no visible difference. Energy mode integrates in the chamber and stops at the set dose. Dose setting for membrane crosslinking

What dose should a nylon or nitrocellulose membrane get?

120 mJ/cm² is the usual starting point and the default on most auto-crosslink routines. If signal is weak you can try 150 to 200 mJ/cm², but over-crosslinking binds nucleic acids so tightly that probe access suffers. Run a gradient on one membrane lot rather than changing one variable at a time across weeks. Crosslinker dose setup guide

Do I need a transilluminator as well as a crosslinker?

They do different jobs. A crosslinker is a closed chamber irradiating from above to deliver a controlled dose to a membrane. A transilluminator is an open bench surface lit from below for viewing, cutting and imaging gels. Using one for the other's purpose gives an uncontrolled dose and poor reproducibility. Transilluminator range

How do I know when the tubes need replacing?

Not by whether they still glow. Record the seconds needed for a given dose when the instrument is new, and compare annually: if it takes more than about twice as long, replace the tubes. A radiometer at the chamber centre plane confirms it. Comparing seven crosslinker models

Fluorescent leak detection

How much dye goes into the system?

Oil-soluble dye at 0.5 to 2.5 ml per litre of oil, run for 5 to 10 minutes, inspected with a 365 nm lamp; no flushing is needed afterwards because oil performance is unaffected. Water-based dyes dilute from 1000:1 to 500:1. Leak detection guide with 18 field questions

Can oil-based and water-based dyes be mixed?

No. An oil-soluble dye added to a water circuit separates and fails to travel; a water-based formula cannot disperse evenly in an oil circuit. Identify the medium first, then pick the dye family. Dye and lamp range

How long does a micro-leak take to show?

It depends on leak rate and circulation: an obvious leak shows in minutes, while a very small one may need hours or a couple of days of running for the dye to seep to the surface and build a visible concentration. Extend the run time or clean and recheck before raising the dose. Making early leak detection effective

How do I avoid mistaking old residue for a new leak?

Clean thoroughly around the original leak after the test, then recheck. Residual fluorescence from a previous inspection is one of the more common false positives, and it is entirely avoidable with a wipe and a second look. Leak detection in preventive maintenance

Biofilm and cleanroom hygiene

Can a UV lamp replace ATP swabs or plate counts?

No, and it should not be sold as if it could. Fluorescence shows where to sample: sweep the surface, mark the fluorescent spots, then take targeted ATP or plate samples at those spots. The regulatory criterion stays with the swab and the plate. Biofilm detection with PMA-qPCR and fluorescence

What do green and blue fluorescence mean?

Under UV-A, green usually comes from protein-containing contamination such as biofilm and active bacteria, which does not wipe away cleanly. Blue usually comes from dust and fibres, which do. Distinguishing them prevents both a shutdown over loose dust and a real biofilm being wiped over as if it were dirt. Biofilm detection lamp range

Culture came back negative. Do I still need to inspect?

Yes. Under nutrient limitation or disinfectant stress, bacteria can enter a dormant or viable-but-non-culturable state: alive, but not forming colonies on the plate. Fluorescence reads extracellular protein, which is independent of culturability, so it covers that blind spot. Why pharmaceutical plants have to monitor biofilm

Does it have to be dark, and how fast can a patrol go?

A darkroom helps but is not mandatory: dim the lights, use a hood, or work inside equipment where ambient light is low. With the lamp head 10 to 30 cm from the surface and one-third overlap you cover roughly 2 to 4 m² per minute, so a 500 m² workshop focuses on equipment surfaces, welds and drains in one to two hours. Biofilm inspection in pharmaceutical cleanrooms

Laboratory irradiation and phototoxicity testing

What does an OECD 432 phototoxicity test require from the light source?

A defined UVA dose delivered consistently, with the spectrum and irradiance documented. That means a source with stable output, a means of measuring it, and a record of the dose actually received by the plate rather than the nominal setting. 18 FAQs on cell phototoxicity irradiation

How is UVA dose calculated for a 96-well plate?

Measure irradiance at the plate plane in mW/cm², multiply by exposure time in seconds to get mJ/cm², and correct for any difference between the calibration plane and the wells. Published protocols usually state the target dose; the instrument should be able to reproduce it. Animal phototoxicity testing conditions

Why does the same protocol give different results in two labs?

Almost always geometry and dose: distance, beam uniformity, plate type and the age of the lamps. Two setups that both claim the same irradiance can deliver different doses at the well surface. Record the measured value, not the dial setting. Phototoxicity assessment of sunscreens

How do I choose between an irradiation box and a lamp panel?

By sample format and how much control the protocol needs. A closed box gives defined geometry and repeatable dose across a plate; an open panel is more flexible for odd sample shapes but needs the distance and uniformity established by the user. Irradiation system range

Did not find the question?

Send the application, the acceptance document and the measurement conditions, and the answer comes back against your clause rather than a general figure. Contact Xinysun Instruments, or start from the product range and the company news and application notes.